Krohn-Hite
|
Model PCS-2/B
current-to-voltage calibrator
or transconductance amplifier
|
 |
GENERAL
DESCRIPTION
The Model PCS-2/B is a current-to-voltage calibrator or transconductance
amplifier. The current output is directly proportional to the input
voltages from a high accuracy AC or DC voltage standard. The input voltage
to the unit can be AC or DC up to 1kHz. The ratios of 10 volts to output
is: 1:1, 1:0.1, 1:0.01, 1:0.001 and 1:0.0001. The PCS-2/B is laboratory
calibrated with instruments and load resistors traceable to N.I.S.T.
(National Institute of Standards & Technology). The unit is also equipped
for rack mounting in a standard 19" rack.
Recommended
for DC – Models 511 and 521
The PCS-2B may also be used with AC and DC voltage sources (calibrators)
of other manufacturers.
APPLICATIONS
Applications for
the Model PCS-2B include: Direct Calibration of Watt Meters, Meter Shunts,
AC and DC Current Meters, I to E Transducers, DVMs, DMMs, Relay Coil
Current Testing.
SPECIFICATIONS
INPUT
Input Voltage:
0V to ±11Vdc or 0V to 11Vac (rms).
Input Impedance: 10k ohms (all ranges).
AC Bandwidth: 40Hz to 1kHz.
OUTPUT
|
Ranges |
Resolution |
Compliance
Voltage |
Load
Regulation |
Distortion*
(50Hz to 1kHz) |
|
10A |
10µA |
1.5Vdc
or peak ac |
±0.015% |
<0.5% |
|
1A |
1µA |
5Vdc or
peak ac |
±0.015% |
<0.1% |
|
100mA |
100nA |
5Vdc or
peak ac |
±0.01% |
<0.1% |
|
10mA |
10nA |
5Vdc or
peak ac |
±0.01% |
<0.1% |
|
1mA |
1nA |
5Vdc or
peak ac |
±0.01% |
<0.1% |
|
*
Distortion: Added to the distortion of the AC voltage calibrator. |
Accuracy: DC, ±(0.01% of
setting + 0.01% of range + 10µA);
AC, ±(0.05% of setting + 0.01% of range + 20µA).
Line Regulation: ±0.0005% for ±10% line change.
Stability: DC, 0.001% of range; AC, 0.005% of range.
Isolation: Floating output; limited only by the isolation of the
driving voltage source. Ripple and Noise, rms, non-additive, in a
band-pass 0.1Hz to 100kHz: 1mA range, 1.5µA; 10mA range, 3.0µA; 100mA
range, 40µA; 1A range, 80µA; 10A range, 375µA.
PROGRAMMING
(IEEE-488.1)
The 5 ranges are GPIB
programmable as well as controlled locally with front panel controls. A
dip-switch, located on the rear panel, is used to set the instruments bus
address.
GENERAL
Power
Requirements: 117Vac, 60Hz, 150 watts or 220Vac, 50Hz.
Temperature Coefficient: 10A and 1A ranges, 50ppm/°C;
100mA, 10mA and 1mA ranges, 10ppm/°C.
Operating Temperature: –10°C to 40°C.
Dimensions: 5.22" (12.25cm) high, 17" (43.18cm) wide, 15"
(38.1cm) deep.
Weight: 30 lbs. (13.6kg) net; 34 lbs. (15.4kg) shipping.
Protection: A front panel lamp indicates when the
compliance voltage has exceeded or when a malfunction has occurred.
Certification: Certificate of Compliance traceable to
U.S. N.I.S.T.
Warranty: One full year, includes accuracy and stability.
Extended 1 Year Warranty: Part No. EWPCS2B.
Specifications is subject to change without notice.

|
|